Irian Shafiq, Sharif Hashmani.
Comparison between central corneal thickness measurements obtained with Orbscan II topographer and Ultrasonic Pachymeter
Pak J Ophthalmol Apr 2007;23(2):87-91.
Dow University of Health sciences & Civil Hospital, Karachi
Purpose: To compare the central corneal thickness measurements obtained with Orbscan II scanning slit topographer and ultrasonic pachymeter in eyes undergoing corneal refractive surgery. Material and Methods: This non-interventional, observational comparative study was conducted in Hashmanis Eye Hospital Karachi. A total of 108 eyes evaluated for central corneal thickness measurement with Orbscan II scanning slit topographer and ultrasonic pachymeter that had undergone corneal refractive surgeries (LASIK, LASEK, PRK). The average age of the patients was 29.85 years. Result: The mean central corneal thickness (CCT) was 537.44pm standard deviation (SD) ± 27.31mcm standard error mean (SEM) ±2.63mcm with Orbscan II scanning slit topographer and mean CCT 542.04 pm SD ±26.35pm and SEM ±2.54pm with ultrasonic pachymeter (P = 0.007). Conclusion: The mean Central corneal thickness (CCT) from both devices was found not statistically significantly different from each other.
Category: Eye
Keywords: Cornea. Pachymeter. Ultrasonography.
Click here for Full Text Article