Muhammad Saqlain Tahir, Ahmad Adnan, Quratulain Syed.
chromatographic resolution of drug analogues: 3-hydroxy-3-methylglutarylcoenzyme a reductase inhibitors (statins)..
Pak J Pharm Sci Jan ;29(3-S):1043-51.

A high performance liquid chromatographic method for the simultaneous determination both qualitative and quantitative of cholesterol lowering statin drugs in pharmaceutical formulations has been developed. The most important advantage of developed method is that all seven statin drugs can be determined on a single chromatographic system without modification in detection wavelength. An organic modifier addition (25% v/v methanol) in the presence of buffer (20mM ammonium acetate; pH 4.0 adjusted with dilute acetic acid) played a key role in the resolution of statin drugs in gradient elution with acetonitrile. The drugs were separated on a Purospher Star 4.6mm x 25cm, 5mum, C18 column maintained at 25 degrees C with 1mLmin(-1) flow rate using ultra violet detection at 240nm. Good separation (Rs > 2.5) was achieved in a short analysis allowing simultaneous determination of all seven statins. The effect of variation in flow rate, detection wavelength and column oven temperature was also studied. The proposed method was statistically validated in terms of precision, accuracy, linearity, specificity and robustness. The newly developed method proved to be specific, robust and accurate for the quantification of seven statins in commercial pharmaceutical formulations.